Abstract
In this paper the main influence quantities of optical density of Agfa Personal Monitoring film which have a significant component of uncertainty of dose measurement are treated and discussed. The energy and angle dependence for photographic dosemeter were studied for different energies and angles. The aim of this paper is to present and compare the impact on the uncertainty of dose measurement of these influence quantities, in order to review the accuracy of photographic dosemeter for fulfilling the newest EU Technical Recommendations.
Keywords
Radiation protection, Dose assessment, Photographic dosemeter, Reference radiation.
Citation
M. CHERESTES, C. CHERESTES, M. V. PARASCHIVA, L. M. CONSTANTINESCU, C. BÂRCA, Influence quantities of optical density for Agfa personal monitoring film used in personal dosimetry, Optoelectronics and Advanced Materials - Rapid Communications, 4, 6, June 2010, pp.867-870 (2010).
Submitted at: April 15, 2010
Accepted at: June 16, 2010